| Test | Result | Time | Details / log |
|---|---|---|---|
| MCU + firmware | PASS | 08:49:58 |
[mcu] version: 5ac7c301
[mcu] board=SMART_MAKER_1_X;cpu=STM32F401;conn=UART;device=C235;x=SMART_TMC2209_V2;y=SMART_TMC2209_V2;z=SMART_TMC2209_V2;e0=SMART_TMC2209_V2;e1=SMART_TMC2209_V2;mx=42BYGH3023-B-22DH;my=42BYGH3023-B-22DH;mz=42BYGH2216-B-24DH;piezo=YN15T-85A1-W100_15-2P;pzb=Smart_Z-Offset;pzbv=0.2;mains=220V;bedw=300;ssr=SSR10DA;lot=20260616;uid=61C229
[mcu] comment: -
[mcu SmartPiOne] version: 5ac7c301 (host SmartPi One)
MCU_UID=270051000C51353234323830
|
| Home X | PASS | 08:50:19 |
Run Current: 0.55A Hold Current: 0.55A
YUMI_SENSORLESS_HOME X: home base...
Run Current: 1.20A Hold Current: 1.09A
tap 1: pos=256.5697 gap=1.4303 (chauffe, ignore)
tap 2: pos=255.9969 gap=2.0031 (1/4)
tap 3: pos=255.9969 gap=2.0031 (2/4)
tap 4: pos=255.9969 gap=2.0031 (3/4)
tap 5: pos=255.9969 gap=2.0031 fenetre=0.0000/0.0500
YUMI_SENSORLESS_HOME X OK: 4 taps valides (0 rejetes) -> moyenne=255.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=256.0000
|
| Home Y | PASS | 08:50:49 |
Run Current: 0.55A Hold Current: 0.55A
YUMI_SENSORLESS_HOME Y: home base...
Run Current: 1.20A Hold Current: 1.20A
tap 1: pos=-30.3869 gap=1.6131 (chauffe, ignore)
tap 2: pos=-29.9969 gap=2.0031 (1/4)
tap 3: pos=-29.9969 gap=2.0031 (2/4)
tap 4: pos=-29.9847 gap=2.0153 (3/4)
tap 5: pos=-30.0091 gap=1.9909 fenetre=0.0244/0.0500
YUMI_SENSORLESS_HOME Y OK: 4 taps valides (0 rejetes) -> moyenne=-29.9969 spread=0.0244mm (tol=0.0500). Zero pose en butee=-30.0000
|
| Motherboard fan | PASS | 08:51:01 | |
| Part fan | PASS | 08:51:04 | |
| Hotend fan | PASS | 08:51:07 | |
| Hotend heat 220°C | PASS | 08:51:10 | |
| Bed heat 60°C | PASS | 08:51:30 | |
| Cutter (feed+cut) | PASS | 08:52:53 |
Run Current: 0.55A Hold Current: 0.55A
YUMI_SENSORLESS_HOME X: home base...
Run Current: 1.20A Hold Current: 1.09A
tap 1: pos=256.5819 gap=1.4181 (chauffe, ignore)
tap 2: pos=255.9969 gap=2.0031 (1/4)
tap 3: pos=255.9969 gap=2.0031 (2/4)
tap 4: pos=255.9969 gap=2.0031 (3/4)
tap 5: pos=255.9969 gap=2.0031 fenetre=0.0000/0.0500
YUMI_SENSORLESS_HOME X OK: 4 taps valides (0 rejetes) -> moyenne=255.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=256.0000
YUMI_SENSORLESS_HOME Y: home base...
Run Current: 1.20A Hold Current: 1.20A
tap 1: pos=-30.1919 gap=1.8081 (chauffe, ignore)
tap 2: pos=-30.1797 gap=1.8203 (1/4)
tap 3: pos=-29.9969 gap=2.0031 (2/4)
tap 4: pos=-29.8141 gap=2.1859 (3/4)
tap 5: pos=-30.1797 gap=1.8203 fenetre=0.3656/0.0500
YUMI_SENSORLESS_HOME Y IMPRECIS: 4 taps valides (0 rejetes) -> moyenne=-30.0426 spread=0.3656mm (tol=0.0500). Zero pose en butee=-30.0000
Extruder 'extruder1' now syncing with ''
Extruder 'extruder0' now syncing with 'extruder'
QC CUTTER: motion sensor YMS-1 a change d'etat (mouvement detecte)
QC CUTTER: extrude 60 + refroidit poop 5s + coupe + retracte 120
Extruder 'extruder0' now syncing with ''
|
| YMS-2 feed to head | PASS | 08:53:03 |
Extruder 'extruder0' now syncing with ''
Extruder 'extruder1' now syncing with ''
Extruder 'extruder1' now syncing with 'extruder'
QC E1_HEAD: motion sensor YMS-2 a change d'etat (mouvement detecte)
QC E1_HEAD: filament a la tete apres 50mm + motion sensor YMS-2 OK
|
| Z tap home + Zmax | PASS | 08:56:05 |
Run Current: 0.55A Hold Current: 0.55A
YUMI_SENSORLESS_HOME X: home base...
Run Current: 1.20A Hold Current: 1.09A
tap 1: pos=256.3869 gap=1.6131 (chauffe, ignore)
tap 2: pos=256.1919 gap=1.8081 (1/4)
tap 3: pos=255.9969 gap=2.0031 (2/4)
tap 4: pos=255.9969 gap=2.0031 (3/4)
tap 5: pos=255.9969 gap=2.0031 fenetre=0.1950/0.0500
YUMI_SENSORLESS_HOME X IMPRECIS: 4 taps valides (0 rejetes) -> moyenne=256.0456 spread=0.1950mm (tol=0.0500). Zero pose en butee=256.0000
YUMI_SENSORLESS_HOME Y: home base...
Run Current: 1.20A Hold Current: 1.20A
tap 1: pos=-30.3747 gap=1.6253 (chauffe, ignore)
tap 2: pos=-29.8141 gap=2.1859 (1/4)
tap 3: pos=-30.1797 gap=1.8203 (2/4)
tap 4: pos=-29.8141 gap=2.1859 (3/4)
tap 5: pos=-29.9969 gap=2.0031 fenetre=0.3656/0.0500
YUMI_SENSORLESS_HOME Y IMPRECIS: 4 taps valides (0 rejetes) -> moyenne=-29.9512 spread=0.3656mm (tol=0.0500). Zero pose en butee=-30.0000
Moving to tap point X=31.0 Y=212.0...
Probing with pressure switch...
probe at 31.003,211.998 is z=156.182500
probe at 31.003,211.998 is z=156.210000
Tap drift (diff=0.0275mm) — reset
probe at 31.003,211.998 is z=156.185000
Tap drift (diff=0.0250mm) — reset
Tap OK (diff=0.0025mm, stable 2/5)
probe at 31.003,211.998 is z=156.207500
Tap OK (diff=0.0000mm, stable 2/5)
Tap OK (diff=0.0025mm, stable 3/5)
Tap OK (diff=0.0000mm, stable 4/5)
Tap OK (diff=0.0000mm, stable 3/5)
Tap OK (diff=0.0000mm, stable 5/5)
VALIDATED: trigger_z=156.2100 -> Z=0 pose 0.1000 au-dessus du tap (compression=0.0000 + z_offset=0.1000)
Maillage OK (config): zero_reference_position (31.0, 212.0) ~ point de tap (31.0, 212.0)
Z calibration complete!
|
| Z tap repeatability | PASS | 09:02:25 |
OK: 3 taps convergents spread=0.0000mm (tol=0.0500) sur 15 taps | taps=0.9075, 255.9000, 255.9000, 255.9000, 256.0150, 255.9050, 255.9375, 255.9075, 255.8975, 255.8975, 255.9075, 255.8925, 255.9050, 255.8925, 255.9125
Moving to tap point X=31.0 Y=212.0...
Probing with pressure switch...
probe at 30.994,211.998 is z=0.870000
probe at 30.994,211.998 is z=0.875000
Tap OK (diff=0.0050mm, stable 2/5)
probe at 30.994,211.998 is z=0.902500
Tap drift (diff=0.0275mm) — reset
Tap OK (diff=0.0000mm, stable 2/5)
probe at 30.994,211.998 is z=0.905000
Tap OK (diff=0.0025mm, stable 3/5)
probe at 30.994,211.998 is z=0.907500
Tap OK (diff=0.0025mm, stable 4/5)
Tap OK (diff=0.0000mm, stable 5/5)
VALIDATED: trigger_z=0.9075 -> Z=0 pose 0.1000 au-dessus du tap (compression=0.0000 + z_offset=0.1000)
Maillage OK (config): zero_reference_position (31.0, 212.0) ~ point de tap (31.0, 212.0)
Z calibration complete!
probe at 30.994,211.998 is z=255.895000
probe at 30.994,211.998 is z=255.897500
Tap OK (diff=0.0025mm, stable 2/5)
probe at 30.994,211.998 is z=255.900000
Tap OK (diff=0.0000mm, stable 4/5)
VALIDATED: trigger_z=255.9000 -> Z=0 pose 0.1000 au-dessus du tap (compression=0.0000 + z_offset=0.1000)
probe at 30.994,211.998 is z=255.902500
probe at 30.994,211.998 is z=255.905000
Tap OK (diff=0.0050mm, stable 5/5)
probe at 30.994,211.998 is z=255.862500
Tap drift (diff=0.0375mm) — reset
Tap drift (diff=0.0400mm) — reset
Tap OK (diff=0.0000mm, stable 3/5)
probe at 30.994,211.998 is z=255.990000
Tap drift (diff=0.0850mm) — reset
probe at 30.994,211.998 is z=255.935000
Tap drift (diff=0.0550mm) — reset
probe at 30.994,211.998 is z=255.992500
Tap drift (diff=0.0575mm) — reset
probe at 30.994,211.998 is z=255.937500
probe at 30.994,211.998 is z=255.987500
Tap drift (diff=0.0500mm) — reset
probe at 30.994,211.998 is z=255.887500
Tap drift (diff=0.1000mm) — reset
|
| Screws tilt | PASS | 09:03:19 |
probe: at 32.495,205.497 bed will contact at z=2.237500
probe: at 32.495,205.497 bed will contact at z=2.407500
Probe samples exceed tolerance. Retrying...
probe: at 32.495,205.497 bed will contact at z=2.402500
probe: at 32.495,205.497 bed will contact at z=2.397500
probe: at 32.495,205.497 bed will contact at z=2.392500
probe: at 206.996,205.497 bed will contact at z=3.615000
probe: at 206.996,205.497 bed will contact at z=3.785000
probe: at 206.996,205.497 bed will contact at z=3.795000
probe: at 206.996,205.497 bed will contact at z=3.642500
probe: at 206.996,205.497 bed will contact at z=3.800000
probe: at 206.996,205.497 bed will contact at z=3.622500
probe: at 206.996,205.497 bed will contact at z=3.792500
probe: at 206.996,205.497 bed will contact at z=3.627500
probe: at 206.996,205.497 bed will contact at z=3.807500
probe: at 206.996,205.497 bed will contact at z=3.617500
probe: at 206.996,205.497 bed will contact at z=3.810000
probe: at 206.996,205.497 bed will contact at z=3.637500
probe: at 206.996,205.497 bed will contact at z=3.787500
probe: at 206.996,205.497 bed will contact at z=3.635000
probe: at 206.996,205.497 bed will contact at z=3.797500
probe: at 206.996,31.995 bed will contact at z=3.902500
probe: at 206.996,31.995 bed will contact at z=4.060000
probe: at 206.996,31.995 bed will contact at z=4.070000
probe: at 206.996,31.995 bed will contact at z=4.055000
probe: at 32.495,31.995 bed will contact at z=2.702500
probe: at 32.495,31.995 bed will contact at z=2.862500
probe: at 32.495,31.995 bed will contact at z=2.830000
probe: at 32.495,31.995 bed will contact at z=2.840000
probe: at 32.495,31.995 bed will contact at z=2.845000
01:20 means 1 full turn and 20 minutes, CW=clockwise, CCW=counter-clockwise
front left screw (base) : x=49.5, y=175.5, z=2.39750
front right screw : x=224.0, y=175.5, z=3.79750 : adjust CW 02:48
rear right screw : x=224.0, y=2.0, z=4.06000 : adjust CW 03:19
rear left screw : x=49.5, y=2.0, z=2.84000 : adjust CW 00:53
|
| QC date | Result | QC pad (MAC) | Received (server) |
|---|---|---|---|
| 2026-06-27T08:49:58.284391 | PASS | 270051000C51353234323830 | 2026-06-29T00:38:50 |
| 2026-06-27T08:46:07.391092 | FAIL | 270051000C51353234323830 | 2026-06-27T08:41:38 |