SmartPi 1
SmartPi 2

4E0067000F51353135323934

PASS Grade A Grade B
Model : C235 · Batch : 20260616
Heated bed : 220V · 300W · SSR SSR10DA
QC pad (MAC) : 0281DE6A3114
Start : 2026-07-02T02:14:39.190795 · End : 2026-07-02T02:26:09.850802 · Duration : 690 s
Received (server) : 2026-09-11T09:23:40
yumi_config : board=SMART_MAKER_1_X cpu=STM32F401 conn=UART device=C235 x=SMART_TMC2209_V2 y=SMART_TMC2209_V2 z=SMART_TMC2209_V2 e0=SMART_TMC2209_V2 e1=SMART_TMC2209_V2 mx=42BYGH3023-B-22DH my=42BYGH3023-B-22DH mz=42BYGH2216-B-24DH piezo=YN15T-85A1-W100_15-2P pzb=Smart_Z-Offset pzbv=0.2 mains=220V bedw=300 ssr=SSR10DA lot=20260616 uid=61C229

Proof photos (packing)

Switch to mobile pipeline mode
TestResultTimeDetails / log
MCU + firmware PASS 02:14:39
[mcu] version: 5ac7c301 [mcu] board=SMART_MAKER_1_X;cpu=STM32F401;conn=UART;device=C235;x=SMART_TMC2209_V2;y=SMART_TMC2209_V2;z=SMART_TMC2209_V2;e0=SMART_TMC2209_V2;e1=SMART_TMC2209_V2;mx=42BYGH3023-B-22DH;my=42BYGH3023-B-22DH;mz=42BYGH2216-B-24DH;piezo=YN15T-85A1-W100_15-2P;pzb=Smart_Z-Offset;pzbv=0.2;mains=220V;bedw=300;ssr=SSR10DA;lot=20260616;uid=61C229 [mcu] comment: - [mcu SmartPiOne] version: 5ac7c301 (host SmartPi One) MCU_UID=4E0067000F51353135323934
Home X PASS 02:15:02
Run Current: 0.55A Hold Current: 0.55A YUMI_SENSORLESS_HOME X: home base... Run Current: 1.20A Hold Current: 1.09A tap 1: pos=256.5697 gap=1.4303 (chauffe, ignore) tap 2: pos=255.9969 gap=2.0031 (1/4) tap 3: pos=255.9969 gap=2.0031 (2/4) tap 4: pos=255.9969 gap=2.0031 (3/4) tap 5: pos=255.9969 gap=2.0031 fenetre=0.0000/0.0500 YUMI_SENSORLESS_HOME X OK: 4 taps valides (0 rejetes) -> moyenne=255.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=256.0000
Home Y PASS 02:15:33
Run Current: 0.55A Hold Current: 0.55A YUMI_SENSORLESS_HOME Y: home base... Run Current: 1.20A Hold Current: 1.20A tap 1: pos=-30.3869 gap=1.6131 (chauffe, ignore) tap 2: pos=-29.9969 gap=2.0031 (1/4) tap 3: pos=-29.9969 gap=2.0031 (2/4) tap 4: pos=-29.9969 gap=2.0031 (3/4) tap 5: pos=-29.9969 gap=2.0031 fenetre=0.0000/0.0500 YUMI_SENSORLESS_HOME Y OK: 4 taps valides (0 rejetes) -> moyenne=-29.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=-30.0000
Motherboard fan PASS 02:15:56
Part fan PASS 02:15:59
Hotend fan PASS 02:16:02
Hotend heat 220°C PASS 02:16:03
Bed heat 60°C PASS 02:16:13
Cutter (feed+cut) PASS 02:17:39
Run Current: 0.55A Hold Current: 0.55A YUMI_SENSORLESS_HOME X: home base... Run Current: 1.20A Hold Current: 1.09A tap 1: pos=256.5697 gap=1.4303 (chauffe, ignore) tap 2: pos=255.9969 gap=2.0031 (1/4) tap 3: pos=255.9969 gap=2.0031 (2/4) tap 4: pos=255.9969 gap=2.0031 (3/4) tap 5: pos=255.9969 gap=2.0031 fenetre=0.0000/0.0500 YUMI_SENSORLESS_HOME X OK: 4 taps valides (0 rejetes) -> moyenne=255.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=256.0000 YUMI_SENSORLESS_HOME Y: home base... Run Current: 1.20A Hold Current: 1.20A tap 1: pos=-30.1919 gap=1.8081 (chauffe, ignore) tap 2: pos=-29.9969 gap=2.0031 (1/4) tap 3: pos=-29.9969 gap=2.0031 (2/4) tap 4: pos=-29.9969 gap=2.0031 (3/4) tap 5: pos=-29.9969 gap=2.0031 fenetre=0.0000/0.0500 YUMI_SENSORLESS_HOME Y OK: 4 taps valides (0 rejetes) -> moyenne=-29.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=-30.0000 Extruder 'extruder1' now syncing with '' Extruder 'extruder0' now syncing with 'extruder' QC CUTTER: motion sensor YMS-1 a change d'etat (mouvement detecte) QC CUTTER: extrude 60 + refroidit poop 5s + coupe + retracte 120 Extruder 'extruder0' now syncing with ''
YMS-2 feed to head PASS 02:17:50
Extruder 'extruder0' now syncing with '' Extruder 'extruder1' now syncing with '' Extruder 'extruder1' now syncing with 'extruder' QC E1_HEAD: motion sensor YMS-2 a change d'etat (mouvement detecte) QC E1_HEAD: filament a la tete apres 50mm + motion sensor YMS-2 OK
Z tap home + Zmax PASS 02:20:43
Run Current: 0.55A Hold Current: 0.55A YUMI_SENSORLESS_HOME X: home base... Run Current: 1.20A Hold Current: 1.09A tap 1 rejete: aucun contact sur 7.0mm (No trigger on probe after full movement) tap 2: pos=256.7281 gap=1.2719 (chauffe, ignore) tap 3: pos=255.9969 gap=2.0031 (1/4) tap 4: pos=255.9969 gap=2.0031 (2/4) tap 5: pos=255.9969 gap=2.0031 (3/4) tap 6: pos=255.9969 gap=2.0031 fenetre=0.0000/0.0500 YUMI_SENSORLESS_HOME X OK: 4 taps valides (1 rejetes) -> moyenne=255.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=256.0000 YUMI_SENSORLESS_HOME Y: home base... Run Current: 1.20A Hold Current: 1.20A tap 1: pos=-30.1919 gap=1.8081 (chauffe, ignore) tap 2: pos=-29.9969 gap=2.0031 (1/4) tap 3: pos=-29.9969 gap=2.0031 (2/4) tap 4: pos=-29.9969 gap=2.0031 (3/4) tap 5: pos=-29.9969 gap=2.0031 fenetre=0.0000/0.0500 YUMI_SENSORLESS_HOME Y OK: 4 taps valides (0 rejetes) -> moyenne=-29.9969 spread=0.0000mm (tol=0.0500). Zero pose en butee=-30.0000 Moving to tap point X=31.0 Y=212.0... Probing with pressure switch... probe at 31.003,211.998 is z=96.530000 Tap OK (diff=0.0000mm, stable 2/5) Tap OK (diff=0.0000mm, stable 3/5) probe at 31.003,211.998 is z=96.532500 Tap OK (diff=0.0025mm, stable 4/5) Tap OK (diff=0.0000mm, stable 5/5) VALIDATED: trigger_z=96.5325 -> Z=0 pose 0.1000 au-dessus du tap (compression=0.0000 + z_offset=0.1000) Maillage OK (config): zero_reference_position (31.0, 212.0) ~ point de tap (31.0, 212.0) Z calibration complete!
Z tap repeatability PASS 02:25:20 OK: 3 taps convergents spread=0.0000mm (tol=0.0500) sur 15 taps | taps=0.8900, 255.9025, 255.9025, 255.8125, 255.9000, 255.9000, 255.9025, 255.9000, 255.9000, 255.8975, 255.9000, 255.9025, 255.9000, 255.9000, 255.9000
Moving to tap point X=31.0 Y=212.0... Probing with pressure switch... probe at 30.994,211.998 is z=0.900000 probe at 30.994,211.998 is z=0.902500 Tap OK (diff=0.0025mm, stable 2/5) probe at 30.994,211.998 is z=0.890000 Tap drift (diff=0.0125mm) — reset Tap OK (diff=0.0000mm, stable 2/5) Tap OK (diff=0.0000mm, stable 3/5) Tap OK (diff=0.0000mm, stable 4/5) Tap OK (diff=0.0000mm, stable 5/5) VALIDATED: trigger_z=0.8900 -> Z=0 pose 0.1000 au-dessus du tap (compression=0.0000 + z_offset=0.1000) Maillage OK (config): zero_reference_position (31.0, 212.0) ~ point de tap (31.0, 212.0) Z calibration complete! probe at 30.994,211.998 is z=255.900000 probe at 30.994,211.998 is z=255.902500 VALIDATED: trigger_z=255.9025 -> Z=0 pose 0.1000 au-dessus du tap (compression=0.0000 + z_offset=0.1000) Tap OK (diff=0.0025mm, stable 3/5) probe at 30.994,211.998 is z=255.925000 Tap drift (diff=0.0225mm) — reset probe at 30.994,211.998 is z=255.927500 probe at 30.994,211.998 is z=255.905000 Tap drift (diff=0.0200mm) — reset probe at 30.994,211.998 is z=255.907500 Tap drift (diff=0.0175mm) — reset probe at 30.994,211.998 is z=255.932500 Tap OK (diff=0.0050mm, stable 3/5) probe at 30.994,211.998 is z=255.942500 Tap drift (diff=0.0100mm) — reset probe at 30.994,211.998 is z=258.520000 Tap drift (diff=2.5775mm) — reset Tap drift (diff=2.6150mm) — reset probe at 30.994,211.998 is z=255.895000 Tap drift (diff=0.0300mm) — reset probe at 30.994,211.998 is z=255.887500 Tap OK (diff=0.0075mm, stable 2/5) probe at 30.994,211.998 is z=255.852500 Tap drift (diff=0.0350mm) — reset probe at 30.994,211.998 is z=255.817500 probe at 30.994,211.998 is z=255.815000
Screws tilt PASS 02:26:09
probe: at 32.495,205.497 bed will contact at z=1.200000 probe: at 32.495,205.497 bed will contact at z=1.230000 probe: at 32.495,205.497 bed will contact at z=1.170000 Probe samples exceed tolerance. Retrying... probe: at 32.495,205.497 bed will contact at z=1.290000 probe: at 32.495,205.497 bed will contact at z=1.240000 probe: at 32.495,205.497 bed will contact at z=1.175000 probe: at 32.495,205.497 bed will contact at z=1.235000 probe: at 32.495,205.497 bed will contact at z=1.232500 probe: at 32.495,205.497 bed will contact at z=1.245000 probe: at 206.996,205.497 bed will contact at z=2.897500 probe: at 206.996,205.497 bed will contact at z=2.917500 probe: at 206.996,205.497 bed will contact at z=2.927500 probe: at 206.996,31.995 bed will contact at z=2.200000 probe: at 206.996,31.995 bed will contact at z=2.207500 probe: at 206.996,31.995 bed will contact at z=2.220000 probe: at 32.495,31.995 bed will contact at z=0.822500 probe: at 32.495,31.995 bed will contact at z=0.757500 probe: at 32.495,31.995 bed will contact at z=0.745000 probe: at 32.495,31.995 bed will contact at z=0.877500 probe: at 32.495,31.995 bed will contact at z=0.785000 probe: at 32.495,31.995 bed will contact at z=0.720000 probe: at 32.495,31.995 bed will contact at z=0.722500 probe: at 32.495,31.995 bed will contact at z=0.800000 probe: at 32.495,31.995 bed will contact at z=0.815000 probe: at 32.495,31.995 bed will contact at z=0.787500 probe: at 32.495,31.995 bed will contact at z=0.865000 probe: at 32.495,31.995 bed will contact at z=0.775000 probe: at 32.495,31.995 bed will contact at z=0.777500 probe: at 32.495,31.995 bed will contact at z=0.770000 01:20 means 1 full turn and 20 minutes, CW=clockwise, CCW=counter-clockwise front left screw (base) : x=49.5, y=175.5, z=1.23250 front right screw : x=224.0, y=175.5, z=2.91750 : adjust CW 03:22 rear right screw : x=224.0, y=2.0, z=2.20750 : adjust CW 01:57 rear left screw : x=49.5, y=2.0, z=0.77500 : adjust CCW 00:55

History — 3 reports for this machine

QC dateResultQC pad (MAC)Received (server)
2026-07-02T02:14:39.190795 PASS 4E0067000F51353135323934 2026-09-11T09:23:40
2026-06-26T08:33:46.112606 FAIL 4E0067000F51353135323934 2026-06-26T08:29:23
2026-06-26T08:32:07.197857 FAIL 4E0067000F51353135323934 2026-06-26T08:28:41